Electrically heated sample tube
AST 050, AST 100, AST 150
AST样品管
样品管根据不同侧工艺需求采用不同材料制成。可以通电加热,和ASP探头一起工作。AST标准长度为0.5米,1米和1.5米,也可根据需求特制。
Application
The electrically heated Ankersmid sample probe tube
AST 050-100-150-200 is used in extractive sampling systems
to avoid cooling and condensation of the sample.
Condensation can result in blockage of the probe in
combination with dust. Sample gases can be absorbed if
they are cooled and come in contact with condensate.
An extra stainless filter can be mounted on the tip of the tube in case
Of very high dust level ( + 10 g/m³)
Description
This electrical heated sample tube is available in 3 standard lengths: 0.5 , 1 and 1.5 meter.
The standard flange is like all Ankersmid probes a DN65 PN6.
Optional we can deliver adapter flanges for most common process connections.
On the tip of this heated tube, there is possibility to fix a non heating sample probe or pre-filter.
Inside the tube is a PT 100 (max300 °C) or a thermocouple Fe-CuNi integrated. In combination with our digital controller we have a free programmable setpoint and alarm..
In option we have a 2-way Modbus/ RS485 communication, that combines all Ankersmid controllers, so that digital communication with the control room is possible
• Different lengths available
• Completely heated sample tube
• Digital controller
• Easy mounting
• Optional: Modbus/RS485

|
Series ASP Type
|
AST050
|
AST100
|
AST150
|
AST200
|
|
Temperature sensor
|
Pt 100
|
Pt 100
|
Pt 100
|
Pt 100
|
|
Temperature controller
|
Digital, option ,RS485 Modbus
|
|
Probe tube length L1
|
500mm
|
1000mm
|
1500mm
|
2000mm
|
|
Sample temperature max.
|
500°C
|
500°C
|
450°C
|
400°C
|
|
Operating temperature max.
|
200°C
|
200°C
|
200°C
|
200°C
|
|
Pre filter length 500 mm, ø60 mm, filterporosity 2 µm.
|
Option
|
|
Sample gas inlet connection
|
G3/4"i
|
|
Dust loading
|
max. 2 g/m3
|
|
Probe tube volume
|
200ml/m
|
|
Sample pressure max.
|
5 bar g
|
|
Ambient temperature
|
-20 °C to +70 °C
|
|
Storage temperature
|
-30 °C to +70 °C
|
|
Ready for operation
|
After 1hr.
|
|
Power supply
|
230VAC, 600W (115VAC optional)
|
800W
|
1200W
|
1600W
|
|
Electrical connection
|
2 x 2.5 mm² + 2.5 mm²
|
|
Electrical standard
|
EN 61010, EN60519-1
|
|
Degree of protection
|
IP54 EN 60529
|
|
Mounting flange
|
DN65 PN6
|
|
Material of parts in contact with the sample
|
Stainless steel 316 L
|
|
Controller
|
Microprocessor temperature controller with alarm and optional RS485 or Modbus
|
|
|
|
|
|
|
|
|
|