CYANIDE
Cyanide is used in industrial applications ranging from organic synthesis in a chemical plant to precious metal plating. Due to high toxicity, HCN is also a priority pollutant which must be measured on-line to meet federal and state permitted release levels.
A typical application in an HCN production plant would be the measurement of cyanide in effluent waters before chlorination in a cyanide destruction process, and another application would be monitoring low levels of cyanide in waste recovery just prior to final effluent release. In both applications the objectives are:
TYTRONICS'® SENTINEL
Tytronics’ Sentinel is an on-line colorimetric analyzer. The instrument uses a batch sampling technique that allows the use of 0.25" tubing for sample lines, so small amounts of solids will not affect instrument performance. Other advantages of batch sampling are:
-
analytical results are not affected by sample and reagent flows;
-
analyses are only performed as frequently as required, minimizing reagent consumption.
METHOD
This colorimetric analysis uses 2 ml each of reagent 1, reagent 2 and reagent 3.
|
Method:
|
Colorimetric
|
|
Measurement range:
|
0-100 ppb
|
|
Sample size:
|
30 ml
|
|
Reagent 1:
|
phosphate buffer/ chloramine T
|
|
Reagent 2:
|
4-picoline, sodium hydroxide/EDTA
|
|
Reagent 3:
|
potassium carbonate/ methanol/3-methyl-1-phenyl-2-pyrazoline-5-one
|
|
Volume/analysis each reagent
|
2 ml
|
|
Measurement Wavelength
|
630 nm
|
|
Reference Wavelength
|
810 nm
|
|
Analysis Time
|
15minutes
|
Enhanced Features of Tytronics’ Sentinel:
-
Programmable from 1 to 10 methods for a single analysis compound;
-
Up to 3 different calibration methods;
-
Eight relay outputs, multi-programmable on each relay;
-
Alarm functions, including sample flow, reagent level detection;
-